SenseAir CO_2 Sensor¶
The senseair
sensor platform allows you to use SenseAir CO_2 sensor
(website) with ESPHome.
As the communication with the SenseAir is done using UART, you need
to have an UART bus in your configuration with the rx_pin
connected to the TX pin of the
sensor and the tx_pin
connected to the RX Pin (it’s switched because the
TX/RX labels are from the perspective of the SenseAir sensor). Additionally, you need to set the baud rate to 9600.
# Example configuration entry
uart:
rx_pin: D0
tx_pin: D1
baud_rate: 9600
sensor:
- platform: senseair
co2:
name: "SenseAir CO2 Value"
update_interval: 60s
Configuration variables:¶
co2 (Required): The CO_2 data from the sensor in parts per million (ppm).
All options from Sensor.
update_interval (Optional, Time): The interval to check the sensor. Defaults to
60s
.uart_id (Optional, ID): Manually specify the ID of the UART Component if you want to use multiple UART buses.
id (Optional, ID): Manually specify the ID used for actions.
Note
G+
should be connected to power supply (supported voltage is 4.5 V to 5.25 V), G0
to GND
pin
senseair.background_calibration
Action¶
This action initiates a background calibration on the sensor with the given ID: the current CO2 level will be used as a reference for the 400ppm threshold. Ensure that the sensor is in a stable environment with fresh ambient air, preferably near a window that has already been opened for a sufficient time.
on_...:
then:
- senseair.background_calibration: my_senseair_id
senseair.background_calibration_result
Action¶
This action requests the result of the background calibration procedure from the sensor with the given ID. The value will be printed in ESPHome logs.
Wait at least one sensor lamp cycle after having triggered the background calibration before requesting its result.
on_...:
then:
- senseair.background_calibration_result: my_senseair_id
senseair.abc_get_period
Action¶
This action requests the currently configured ABC interval from the sensor with the given ID. The value will be printed in ESPHome logs.
on_...:
then:
- senseair.abc_get_period: my_senseair_id
senseair.abc_enable
Action¶
This action enables Automatic Baseline Calibration on the sensor with the given ID. ABC will be activated with the default interval of 180 hours.
on_...:
then:
- senseair.abc_enable: my_senseair_id
senseair.abc_disable
Action¶
This action disables Automatic Baseline Calibration on the sensor with the given ID.
on_...:
then:
- senseair.abc_disable: my_senseair_id